Agilent 1260 Infinity Variable Wavelength HPLC Detector G1314F (VWD)

The Agilent 1260 Infinity Variable Wavelength Detector (VWD) is highly sensitive for HPLC and RRLC analyses. It uses time-programmable wavelength switching so you can optimize the detection sensitivity and selectivity for your specific applications. The latest electronics and refined optics have lowered detector noise levels to below 2.5 µAU. The lowest baseline drift (< 1 x 10-4 AU/h) means you can precisely quantify trace level compounds in samples.
High sensitivity with lowest baseline drift achieved by significant hardware improvements over the most successful predecessor Agilent 1200 Series VWD.Electronic temperature control (ETC) - maximum baseline stability and practical sensitivity under fluctuating ambient temperature and humidity conditions.
Up to 100% resolution gain in fast LC by 80 Hz data acquisition rate.Wide linear range - for reliable, simultaneous quantification of primary compounds, by-products and impurities.
Stop-flow wavelength scanning for quick wavelength optimization.Radio frequency identification (RFID) technology for flow cells and lamps - a new level of data traceability.
Automatic wavelength verification by built-in holmium oxide filter.Extensive diagnostics, error detection and display with Instant Pilot controller and Agilent Lab Advisor software.

Specifications:

  • Agilent 1260 Infinity Variable Wavelength Detector (G1314F)
  • Detection type Double-beam photometer
  • Light source Deuterium lamp
  • Number of signals 1
  • Maximum data rate 80 Hz
  • Noise < ± 0.25 x 10 -5 AU, at 230 nm*
  • Drift 1 x 10-4AU/hr, at 230 nm
  • Linearity >2.5 AU upper limit
  • Wavelength range 190-600 nm
  • Wavelength accuracy ± 1 nm, self-calibration with deuterium lines, verification with holmium oxide filter
  • Slit width 6.5 nm typical over whole wavelength range
  • Time programmable Wavelength, polarity, peak width, lamp on/off.